DESCRIPTION:
Standard Anti-Blok Ventricular Catheter-Reservoirs are designed
with four axial channels at the tip with recessed rectangular molded
holes. The channels promote free flow of CSF and discourage occlusion
of the holes by debris, tissue, choroid plexus or ventricle walls.
The reservoir may be tapped with a 25G or smaller Hueber-point needle
for CSF sampling or injection of antibiotics or contrast media.
.
|
|
A polyester reinforced base extends beyond the reservoir
diameter to allow fixation sutures. A removable nylon plug allows
conversion of the ventriculostomy reservoir to use as a shunt reservoir.
The catheter-reservoirs are manufactured of barium impregnated,
silicone elastomer tubing. A stainless steel stylet is provided.
Anti-Blok Ventricular catheter-reservoirs with fishmouth slits for
endoscopic placement.
|