DESCRIPTION:
The Phoenix CRx Universal Shunt consists of one silicone
elastomer, CRx slit valve positioned in a polysulfone housing, which
is enclosed in a cylindrical silicone elastomer pumping chamber
with attached 90 cm, closed end distal tube.
The valve mechanism inside the polysulfone valve housing
is thimble-shaped with a cross-shaped slit in its distal end. The
mid-portion of the cylindrical silicone body serves as a pumping
chamber.
A Polyester re-inforced silicone suture tab is bonded
to the pumping chamber to permit suture fixation to the pericranium.
|
|
There is an integral polysulfone connector at the
inlet of the valve housing to facilitate connection of the ventricular
catheter. The distal tube may be cut to suit the requirements of
the patient.
A radiopaque pressure designation code of one to three
radiopaque dots, indicating low, medium, or high pressure, appears
on the connector tube over the pumping chamber.
Included with the valve is a disposable blunt tip,
stainless steel luer adaptor to facilitate pre-implant valve preparation.
|
|
The Universal Shunt is a 1/2 unitized non-metal, progressive
ressistance CRx valve and pumping chamber, attached 90 cm vision
closed end distal catheter (catheter may be cut to desired length),
an 8 cm or 11 cm Vision SpiralPort ventricular catheter, stylet,
burrhole catheter support and flushing tube.
Available as Shunt Only or in kits featuring a shunt,
46 or 65 cm passer and sheath or shunt and sheath only.
|