 What is
 Hydrocephalus

 

 

 

 

 

 

PRODUCT CATEGORY:
SUB CATEGORY:

About Diamond | Theory of operations | Performance | Diamond Range | Setup Instructions

 Diamond II : DCV220A | DCS202A | DCV225A | DCS225A | DCS225APL | DCS225ASL | DCS225PS | DCS225SS
 Diamond I : DCV120 | DCS102 | DCV125 | DCS125


     VALVES  
   


PERFORMANCE:

 

Figure 4 describes mean inlet fluid pressure resulting from a continuous infusion of fluid, recorded at several discrete flow rates. This reflects in vitro test results and not intracranial pressure following implantation. The nominal pressure specifications, at any given flow rate, may be found by following the horizontal lines to the left side pressure scale.

 

     
Figure 5 describes mean fluid flow rate through the Diamond valve, measured at constant positive poximal pressures. This graph simulates fluid flow while the patient is in a recumbent position, including flow, which may result from pressures, which may be experienced during REM sleep.  
     
Figure 6 describes mean fluid flow rates through the Diamond Valve, measured at constant negative distal pressures, as measured with a positive head pressure of 80 mm H2O. This graph simulates fluid flow rates, which may result due to gravitational forces created between the proximal inlet and the distal outlet of the shunt system, while the patient is either sitting or standing.  
     

 

 

catalog revision date 7/03


<<
    >>

 

 
 