THE PHOENIX DIAMOND™ VALVE

Application: Diamond Valves and Shunt Systems are designed to Provide one-way flow control for shunting cerebrospinal fluid (CSF) from the Ventricles of the brain.

Materials: Silicone elastomer, polysulfone, polyester, and tantalum.

 

Description:
The valve assembly is comprised of two silicone elastometer, differential pressure valve mechanisms in series, connected by an elastic tube, with an aperture in the side wall, which acts as a flow regulating mechanism. The valve mechanism is contained within an integral silicone elastomer chamber with an integral suture base.

Each differential pressure valve mechanism consists of a silicone elastomer dome with two crossed slits in polysulfone housing. Use of polysulfone eliminates MRI artifact. Diamond Valves and Diamond Valve Shunt Systems Contain no Metal Parts. Each differential pressure valve mechanism opens in response to predetermined differential pressure and closes when differential pressure has dropped below a predetermined level.

The design of the valve mechanisms prevents retrograde flow to the brain. The valve housings incorporate integral proximal and distal connectors for easy catheter connection.

 

The operation of the valve mechanism is described below under "Theory of Operation".

The mechanism are contained within a silicone elastomer chamber, which acts as a pumping reservoir. The reservoir is self sealing when penetrated by a 25 gauge or smaller Huner-point needle or equivalent non-coring needle. This permits removal of CSF samples, as well as injection of contrast media, medication or flushing fluid.

The base of the valve is polyester reinforced silicone elastomer to permit suturing of the valve to the pericranium. Integral Hemmer sleeves extend over the connectors,protecting catheter connections from ligature stress and disconnection.

The valve base and all catheters are radiopaque. The valve is marked with a radiopaque flow direction arrow.

     
catalog revision date 7/03


>>